Measurement Calibration

Canonical Definition

Measurement calibration establishes a known correction or reference for part of a measurement chain, such as microphone frequency response, microphone sensitivity, interface response, input full scale, or absolute sound-pressure level. Each calibration applies only to the stage and conditions it covers.

Calibration Types

CalibrationWhat it establishesWhat it does not establish
Microphone frequency-response fileCorrection versus frequency for the stated microphone, orientation, and fileAbsolute SPL unless sensitivity data is included
Microphone sensitivity or SPL calibrationAbsolute pressure-to-digital-level relationship for the unchanged input chainFlat frequency response or complete standards compliance
Interface loopbackElectrical input/output response, delay, gain, and possible channel mismatchMicrophone, loudspeaker, amplifier-load, or acoustic-field correction
Input full-scale calibrationVoltage corresponding to digital full scale at a defined gain settingWaveform cleanliness outside the tested frequency, level, and load
Relative baselineRepeatable before-and-after comparison with the same chainTraceable absolute units

Microphone File And Orientation

Use the calibration file that matches the microphone serial number and intended incidence angle. A file intended for on-axis or 0-degree use should not be silently reused for a 90-degree orientation. The microphone body, grille, windscreen, holder, and nearby surfaces can also influence high-frequency response.

Some USB microphone files include a sensitivity value that allows absolute SPL. Others only provide frequency correction. Read the file and manufacturer instructions instead of assuming.

Absolute SPL Calibration

Absolute sound-pressure measurement requires a known sensitivity reference. That can come from a microphone calibration file containing sensitivity, an acoustic calibrator, or a reference sound level meter used under a documented procedure. Changing operating-system input volume, interface gain, preamp gain, or microphone sensitivity after calibration invalidates the absolute reference.

Calibration is not certification.

IEC 61672 defines complete sound-level-meter instrument performance, including frequency weighting, time weighting, environmental behaviour, tolerances, and measurement uncertainty. Software plus a microphone file should not be advertised as a class-compliant meter unless the entire configured instrument chain has been evaluated accordingly.

Interface Loopback

  1. Connect an output to an input using a safe level and appropriate cable or attenuation.
  2. Confirm neither output nor input clips.
  3. Measure response, delay, channel balance, and distortion over the required range.
  4. Create or apply the correction only to measurements that use the same interface path and settings.
  5. Save the raw loopback measurement and record sample rate, gain, and cable path.

A loopback may expose interface rolloff or clock and latency behaviour, but it cannot correct the acoustic device under test.

Calibration Boost And Noise Floor

A calibration curve that adds large gain where the sensor or interface is weak also raises the displayed noise floor in that region. Do not mistake correction-amplified noise for real acoustic output. Limit or question extreme extrapolation beyond the calibration file's valid range.

Calibration Record